Atlas FDA

Philips IntelliVue Information Center iX (release A.00, A.01, and A.02) are impacted by this issue: 866023 IntelliVue In

FDA device recall Z-0374-2016 · posted 2015-12-04 · Terminated

Recall details

Recalling firm
Philips Medical Systems, Inc.
Reason for recall
Sp02 and/or Non Invasive Blood Pressure (NBP) alarms may become disabled without visual notification
Action taken
The firm, Philips Healthcare, issued a "URGENT-Field Safety Notice" on 12/19/14 to inform customers of the issue, identify details of the units affected, instructs on actions to be taken by the customer and identify what action Philips plans to take to remedy the issue. Philips will provide a software correction free of charge. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Field Safety Notice-To prevent this issue from occurring, customers/users should: 1) Disable the Change Bed Label field in Patient Management. This is accomplished by changing the feature from Optional to Hidden 2) All patient transfers must be performed using the Transfer key in the Manage Patient application. Remember that the most reliable method of patient monitoring combines close personal surveillance with correct operation of monitoring equipment. If you need any further information or support concerning this issue, please contact your local Philips Healthcare Customer Service representative or the Customer Care Solutions Center at (800) 722-9377 with questions or concerns about this Device Correction.
Root cause
Software Design Change
Status
Terminated
Product code
MHX
Quantity affected
10,013 units
Distribution
Worldwide Distribution: US (nationwide) and countries: AUSTRALIA, AUSTRIA, BELGIUM, BULGARIA, CANADA, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, HUNGARY, ICELAND, INDIA, IRAQ, IRELAND, ISRAEL, ITALY,
Date initiated
2014-12-19
Date posted
2015-12-04
Date terminated
2020-09-21
Location
Andover, MA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
M3290B INTELLIVUE INFORMATION CENTER SOFTWARE, RELEASE A.0 (K102495)Philips Medical Systems2010-09-30