Philips IntelliSpace PACS 4.4 image management systems performs digital image processing, measurement and communication

FDA device recall Z-2252-2014 · posted 2014-08-16 · Terminated

Recall details

Recalling firm
Philips Healthcare Informatics, Inc.
Reason for recall
A problem was detected in the Philips IntelliSpace PAC 4.4 software where if patient information is changed in the DICOM header and the study is exported before the image is updated in the PACS, the unchanged data may export causing the potential for diagnosis on incomplete data.
Action taken
Urgent Medical Device Correction was sent to all affected sites on July 24, 2014 by certified mail.
Root cause
Software design
Status
Terminated
Product code
LLZ
Quantity affected
109 units
Distribution
US and Australia, Belgium, Canada, Egypt, France, Germany, Netherlands, Oman, Saudi Arabia, Spain, Switzerland, United Arab Emirates, United Kingdom.
Date initiated
2014-07-24
Date posted
2014-08-16
Date terminated
2016-02-11
Location
Foster City, CA