Atlas FDA

Philips Integris Allura Flat Detector 9C Angiographic X-ray system. This system with Velara Generator can be identified

FDA device recall Z-2477-2010 · posted 2010-09-22 · Terminated

Recall details

Recalling firm
Philips Medical Systems North America Co. Phillips
Reason for recall
A high voltage reliability issue resulting in the loss of the x-ray generation and this can cause an intermittent stop of fluoro acquisition and exposure during an examination without warning and hence no image on the display.
Action taken
The firm, Philips, sent "URGENT-Device Correction" letters dated November 10, 2008, to their customers and informed them about the reliability issue in the Allura XPER and Integris X-ray systems with the High Voltage Velera (HV) Generator. This reliability issue can lead to a system error causing intermittent loss of fluro and/or exposure acquisition, potentially resulting in the loss of the X-ray image on the display. The customers are advised if they experience High Voltage Generator reliability issue and the HV Generator stops working generating an O3EA error, a cold restart is required and reboot the system when the X-Ray image disappears. The firm will implement multiple mandatory Field Change Orders (FCOs) to solve the Generator reliability by updating the software for each X-ray system. The customers could contact Philips Care Center at 800-722-9377, #5, #4, #1 for any questions regarding this recall or 425-487-7469.
Root cause
Other
Status
Terminated
Product code
IZI
Quantity affected
269 units in US
Distribution
Worldwide distribution.
Date initiated
2008-10-23
Date posted
2010-09-22
Date terminated
2011-08-10
Location
Bothell, WA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
PHILIPS INTEGRIS ALLURA FLAT DETECTOR RELEASE 1.0 (K022899)Philips Medical Systems North America Co.2002-11-22