Philips HeartStart XL+ Defibrillator/Monitor Model number: 861290 The HeartStart XL+ is intended for use in a hospital s

FDA device recall Z-0871-2013 · posted 2013-02-21 · Terminated

Recall details

Recalling firm
Philips Healthcare Inc.
Reason for recall
Philips HeartStart XL+ Defibrillator/Monitor may become locked out of clinical use
Action taken
Philips Healthcare sent an Urgent Medical Device Recall letter dated January 2013 on January 28, 2013. The letter identified the product, the problem, and the action to be taken by the customer. Customers were advised if the Philips HeartStart XL+ Defibrillator/Monitor is first powered on and then there is disconnection or an interruption in AC power, the device may be locked out of clinical mode until the user enters the service mode password and performs an Operational Check. Users are provided with the Safety Notice that provides information on how to identify affected devices and instructions on actions to be taken. Customers were advised to follow the ACTION TO BE TAKEN BY CUSTOMER / USER section of the notice. To correct this issue Philips will provide a software upgrade to customers with affected units free of charge. If you need any further information or support concerning this issue, please contact your local Philips representative at 1-800-722-9377. For questions regarding this recall call 978-687-1501.
Root cause
Software design
Status
Terminated
Product code
MKJ
Quantity affected
3785 units
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to ALGERIA, ANGOLA, AUSTRALIA, AUSTRIA, BAHRAIN, BELGIUM, BULGARIA, COLOMBIA, CROATIA (LOCAL NAME HYVATSKA), CYPRUS, CZECH REPUBLIC, DENMA
Date initiated
2013-01-28
Date posted
2013-02-21
Date terminated
2020-09-03
Location
Andover, MA