Philips HeartStart Vehicle Wall Mount, Model M5528A with Philips HeartStart DC Power Module, Model M5529A. The HeartStar
FDA device recall Z-0562-2010 · posted 2009-12-23 · Terminated
Recall details
- Recalling firm
- Philips Healthcare Inc.
- Reason for recall
- HeartStart Vehicle Wall Mount (Model M5528A) with DC Power Module (Model M5529A)--Incorrect wiring during installation by user may result in a smoke or fire hazard.
- Action taken
- Philips Healthcare, Inc. notified customers by an "URGENT - Voluntary Medical Device Correction" letter, plus revised Instructions For Use (IFU) for the Vehicle Wall Mount accessory and the DC Power Module accessory of the HeartStart MRx which includes a revised description of correct installation and verification of performance. Customers were asked to follow the Actions in the customer/user section of the Urgent - Voluntary Medical Device Correction Notification. In this section, customers were given the instructions to correct their devices. Distribution of the Urgent - Voluntary Medical Device Correction Notification along with the IFU will be managed by Philips representatives in each affected geography. Direct questions or concerns about the recall to your local Philips representative.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- MKJ
- Quantity affected
- 3,265 units
- Distribution
- Class II Recall -- Worldwide Distribution -- United States, Canada, Australia, Brazil, China, France, Germany, Hong Kong, Iceland, India, Indonesia, Italy, Japan, Jordan, Kenya, Libyan Arab Jamahiriya
- Date initiated
- 2009-11-12
- Date posted
- 2009-12-23
- Date terminated
- 2012-01-25
- Location
- Andover, MA