Philips HeartStart MRx Monitor/Defibrillators Models: M3535A and M3536A
FDA device recall Z-0204-2016 · posted 2015-11-03 · Terminated
Recall details
- Recalling firm
- Philips Medical Systems, Inc.
- Reason for recall
- 1. Device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade sooner than intended and 2. While connected to AC or DC power and with no battery installed or the battery installed has a charge level of less than 10%, the Ready for Use (RFU) indicator will not provide the expected low battery indication
- Action taken
- Philips Healthcare sent an " Urgent Field Safety Notice " dated November 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. A hardware upgrade will be provided free of charge for all units affected by this issue. A Philips Healthcare representative will contact customers with affected devices to arrange for installation of the upgrade. Philips is asking customers to follow the Action to be Taken by Customer/User section of the Urgent Medical Device Correction notification/Field Safety Notice: A Philips authorized service provider will service the device and ensure it is operating within specification. Prior to receiving service, you can continue to use your MRx provided that the following steps are followed. These steps will prevent further accelerated degradation of the therapy capacitor, and will eliminate the above described behavior for no/low battery conditions. To correct the problem, set the configuration of the MRx to factory defaults in configuration mode. Please contact your local Philips Service Representative for further questions, or call (978) 687-1501.
- Root cause
- Process control
- Status
- Terminated
- Product code
- MKJ
- Quantity affected
- 1553 units
- Distribution
- Worldwide Distribution - USA (nationwide), Australia, Austria, Bahrain, Belgium, Brunei Darussalam, Canada, China, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan,
- Date initiated
- 2014-11-19
- Date posted
- 2015-11-03
- Date terminated
- 2016-08-15
- Location
- Andover, MA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| HEARTSTART MRX, MODELS M3535A OR M3536A (K063375) | Philips Medical Systems | 2007-01-11 |
| HEARTSTART MRX MONITOR/DEFIBRILLATOR (K062233) | Philips Medical Systems North America, Inc. | 2006-11-22 |
| HEARTSTART MRX, INVASIVE PRESSURE & TEMPERATURE OPTIONS (K061707) | Philips Medical Systems | 2006-08-11 |
| HEARTSTART MRX WITH Q-CPR OPTION, MODELS M3535A OR M3536A (K051134) | Philips Medical Systems | 2005-09-08 |
| HEARTSTART MRX, MODEL M3535A (K040404) | Philips Medical Systems, Inc. | 2004-05-04 |
| HEARTSTART MRX, MODEL M3535A (K031187) | Philips Medical Systems | 2003-08-27 |