Atlas FDA

Philips HeartStart MRx Monitor/Defibrillator with CO2/EtCO2 Measurement The HeartStart MRx is for use for the terminatio

FDA device recall Z-0570-2014 · posted 2013-12-24 · Terminated

Recall details

Recalling firm
Philips Medical Systems, Inc.
Reason for recall
MRx defibrillator displays a -?- for EtCO2 and does not display EtCO2 values when patient CO2 level falls below 7.6 mmHg
Action taken
Philips Healthcare issued an Customer Information letter dated February 20, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer with the addendum to the Instructions for Use. If you have a ? in the parameter block and no CO2 waveform on the display, the waveform source is invalid. If invalid, check patient, confirm airway status and examine the FilterLine to make sure it is connected to the device and not kinked or pinched. For more information on the EtCO2 Option see Monitoring Carbon Dioxide chapter in the HeartStart MRx Instructions for Use. Should you have any questions or concerns, please contact your local Philips representative at 1-800-722-9377.
Root cause
Software design
Status
Terminated
Product code
MKJ
Quantity affected
18,529 units
Distribution
US Nationwide and OUS.
Date initiated
2013-12-10
Date posted
2013-12-24
Date terminated
2017-04-18
Location
Andover, MA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
HEARTSTART MRX WITH Q-CPR OPTION, MODELS M3535A OR M3536A (K051134)Philips Medical Systems2005-09-08