Philips Healthcare SureSigns VS2+ Model 863278 SureSigns VS2+ NBP and SpO2 Model 863279 SureSigns VS2+ NBP, SpO2 and Tem

FDA device recall Z-0824-2013 · posted 2013-02-14 · Terminated

Recall details

Recalling firm
Philips Healthcare Inc.
Reason for recall
VS2+ screen may lock up and stop monitoring during patient use
Action taken
Philips issued an Urgent Medical Device Correction letter on January 14, 2013. The field action will take place in two phases: Phase 1 Philips sent customer an Urgent Medical Device Correction Notification/Field Safety Notice. This notification explains the issues and gives the customers instructions on actions to take while they await the hardware correction. Phase 2 Once the hardware correction Philips will contact customers and perform the correction free of charge. Until the correction is available. Affected products may continue to be used while waiting for the correction from Philips if the unit is not used in NBP interval mode. A Philips Healthcare representative will contact you to arrange for the correction of all affected devices. If you need any further information or support concerning this issue, please contact the Philips Customer Care Center at 1-800-722-9377. On 3/14/13, firm stated Initially the fix would be a hardware upgrade but have since changed that to a software update. Customers within the US will be sent a new Field Safety Notice as well as a USB software update package and will be required to notify us upon the upgrade of their affected units. They will be performing the updates themselves
Root cause
Software design
Status
Terminated
Product code
DXN
Quantity affected
3248 units
Distribution
Worldwide Distribution - US Nationwide and the countries of: AUSTRALIA, AUSTRIA, BELGIUM,CANADA, CHILE, DENMARK, FRANCE GERMANY, ICELAND, INDIA, IRELAND,ISRAEL,ITALY, JORDAN, KENYA KUWAIT, LEBANON, MA
Date initiated
2013-01-14
Date posted
2013-02-14
Date terminated
2020-08-27
Location
Andover, MA