Philips DigitalDiagnost Systems with M Cabinet Model: 712025 Stationary radiographic system, digital-Universal RAD Appli
FDA device recall Z-2384-2012 · posted 2012-09-14 · Terminated
Recall details
- Recalling firm
- Philips Healthcare Inc.
- Reason for recall
- Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm.
- Action taken
- Philips Healthcare Imaging issued an Urgent - Field Safety Notice dated August 8, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The unit will be exchanged with a new unit that complies with the IEC requirements. Please reference Field Change Order (FCO) 71200081 when contacting I your local Philips representative. 1-800-722-9377. Customer were instructed to review the information contained in the letter with all members of their staff who need to be aware of the contents of the letter and to maintain a copy of the letter with the equipment instruction for Use. For further assistance, customers should contact their local Phillips representative: Technical Support Line 1-800-722-9377. For questions regarding this recall call 978-687-1501.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- MQB
- Quantity affected
- 4 systems
- Distribution
- Worldwide Distribution - USA including OH and WA and the country of Canada
- Date initiated
- 2012-08-14
- Date posted
- 2012-09-14
- Date terminated
- 2017-01-17
- Location
- Andover, MA