Philips DigitalDiagnost Systems with M Cabinet Model: 712025 Stationary radiographic system, digital-Universal RAD Appli

FDA device recall Z-2384-2012 · posted 2012-09-14 · Terminated

Recall details

Recalling firm
Philips Healthcare Inc.
Reason for recall
Creepage Distance between the two primary phases in the GEO Transformer is 4.5 mm.does not meet (IEC) International standard of 5.5 mm.
Action taken
Philips Healthcare Imaging issued an Urgent - Field Safety Notice dated August 8, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. The unit will be exchanged with a new unit that complies with the IEC requirements. Please reference Field Change Order (FCO) 71200081 when contacting I your local Philips representative. 1-800-722-9377. Customer were instructed to review the information contained in the letter with all members of their staff who need to be aware of the contents of the letter and to maintain a copy of the letter with the equipment instruction for Use. For further assistance, customers should contact their local Phillips representative: Technical Support Line 1-800-722-9377. For questions regarding this recall call 978-687-1501.
Root cause
Device Design
Status
Terminated
Product code
MQB
Quantity affected
4 systems
Distribution
Worldwide Distribution - USA including OH and WA and the country of Canada
Date initiated
2012-08-14
Date posted
2012-09-14
Date terminated
2017-01-17
Location
Andover, MA