Philips Digital Diagnost Software Version R2.0.2 Product Usage: This system is used for making X-ray exposures for diagn

FDA device recall Z-0475-2013 · posted 2012-12-05 · Terminated

Recall details

Recalling firm
Philips Healthcare Inc.
Reason for recall
With Digital Diagnost R2.0.2 a mirrored "R in a circle" is burnt into the image (if mirrored) in the lower right corner of the image. When such an image is mirrored back inside the PACS system, an unmirrored "R in a circle" appears in the lower left corner. This can be mistaken for a "Right patient side" marker, although this can appear on the left patient side.
Action taken
Philips Healthcare initiated a recall on October 25, 2012 via Urgent - Field Safety Notice for the Digital Diagnost X-ray System. The letter identified the affected product, problem and actions to be taken. The letter informs customers that Philips will supply a software update kit to version 2.1.4. For questions please contact your local Philips representative 1-800-722-9377.
Root cause
Software design
Status
Terminated
Product code
MQB
Quantity affected
623 units
Distribution
Worldwide Distribution - USA (nationwide) and the countries of Argentina, Australia, Austria, Belgium, Canada, Chile, Croatia, Czech Republic, Denmark, Egypt, Estonia, Finland, France, Germany, India,
Date initiated
2012-10-30
Date posted
2012-12-05
Date terminated
2017-01-17
Location
Andover, MA