Philips BV Endura, Rel. 2, 718074. Mobile C-Arm X-Ray systems offering Radiographic and Fluoroscopic techniques in a wid

FDA device recall Z-1604-2011 · posted 2011-03-08 · Terminated

Recall details

Recalling firm
Philips Medical Systems North America Co. Phillips
Reason for recall
The fixing strap of the belt holding the CBX Laser to the Image Intensifier may be detached during a procedure when used with the BV Libra and the BV Pulsera/Endura Systems.
Action taken
On 01/06/11, Philips sent an URGENT Device Correction Notice, dated 01/05/2011, to their consignees. The consignees were informed that the fixing strap of the Laser Aiming Device Image Intensifier may inadvertently detach when used with BV Libra, BV Endura Rel. 2, and BV Pulsera Rel 2.3 Mobile X-Ray systems. Specifically the draw latch of the fixing strap holding the CBX Laser to the Image Intensified can be accidentally released during a procedure when used with the above Mobile X-Ray systems. While waiting for Philips to implement their corrective action, the consignees are advised to fix the closed latch body of the draw latch to the metal band using a Tie Wrap. Without securing the latch with a Tie Wrap, the device should not be used. The consignees can contact Philips Call Center at 1-800-722-9377, select 5, and reference FCO 71800035 with any questions about this recall.
Root cause
Device Design
Status
Terminated
Product code
IXL
Quantity affected
10 units
Distribution
Nationwide Distribution -- AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NV, NY, OH, OR, PA, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, a
Date initiated
2010-12-14
Date posted
2011-03-08
Date terminated
2012-02-28
Location
Bothell, WA