Atlas FDA

Philips Bucky DIagnost X-Ray System

FDA device recall Z-0907-2007 · posted 2007-09-25 · Terminated

Recall details

Recalling firm
Philips Medical Systems North America Co. Phillips
Reason for recall
Units do not comply with the requirement to provide automatic beam limitation within +/- 3 degrees of horizontal, vertical, and perpendicular to the plane of the image receptor during radiographic procedures.
Action taken
On 7/23/07 Philips mailed a letter dated July 9, 2007 titled Electronic Product Radiation Warning. The letter advises customers of the issue and asks that customers apply the enclosed cautionary label to the device. The letter also explains that there will be a system modification to be installed. On 8/20/07 the firm sent the same letter, along with a letter dated 8/20/07 to customers thought to have either scrapped their units and the customer had previously indicated they were no longer using the unit. The 8/20/07 letter requested the customer contact Philps if they are currently using the device.
Root cause
Other
Status
Terminated
Product code
KPR
Quantity affected
1,859 units + 65 units added = 1924
Distribution
Units were distributed to 1543 hospitals and medical centers throughout the US. No international distribution from this firm.
Date initiated
2007-07-23
Date posted
2007-09-25
Date terminated
2012-09-17
Location
Bothell, WA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
PHILIPS BUCKY DIAGNOST FAMILY (K945278)Philips Medical Systems North America, Inc.1994-12-14