Philips Brilliance 6, 16, 40, 64, Big Bore, iCT and iCT SP are Computed Tomography X-Ray Systems. The devices are Comput
FDA device recall Z-2035-2012 · posted 2012-07-19 · Terminated
Recall details
- Recalling firm
- Philips Medical Systems (Cleveland) Inc
- Reason for recall
- Philips Healthcare was notified that several of the sixteen bolts securing the brackets which support the carbon table top on various models of their Brilliant CT scanner have been found to be defective and the bolt heads may break off.
- Action taken
- Philips sent a Urgent Medical Device Correction letter dated June 28, 2012, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. Philips Key Market representatives will distribute the letter internationally. Additionally, a Field Service Engineer will visit each site to inspect and correct the issue, and Philips will be tracking the progressof the correction. For further information, Customers were asked to contact their local Philips representative or Philips Healthcare Office. North American and Canadian consignees should contact the Philips Customer Care Solutions Center at 1-800-722-9377
- Root cause
- Process control
- Status
- Terminated
- Product code
- JAK
- Quantity affected
- 524
- Distribution
- Worldwide Distribution -- USA (nationwide) including the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, MJ, NM, NV, NY, OH, OK, OR, PA, PR, TN, T
- Date initiated
- 2012-07-02
- Date posted
- 2012-07-19
- Date terminated
- 2013-12-24
- Location
- Cleveland, OH
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| BRILLIANCE VOLUME (K060937) | Philips Medical Systems (Cleveland), Inc. | 2006-06-05 |
| ACQSIM-MULTISLICE-CT, MODEL 453567042271 (K033357) | Philips Medical Systems (Cleveland), Inc. | 2004-01-09 |
| PHILIPS PLUS COMPUTED TOMOGRAPHY SYSTEM (K033326) | Philips Medical Systems, Inc. | 2003-10-29 |
| MX8000 IDT CT SYSTEM (K012009) | Marconi Medical Systems, Inc. | 2001-09-13 |