Philips Bridge Prep Kit REF: K12-09098B Sterile EO, Rx ONLY
FDA device recall Z-2002-2026 · posted 2026-04-29 · Open, Classified
Recall details
- Recalling firm
- Merit Medical Systems, Inc.
- Reason for recall
- catheter may experience resistance when being advanced over the guidewire
- Action taken
- On April 17, 2026, Merit Medical Issued a Urgent Medical Device Recall notification via: FedEx. Initial notification was issued on April 2, 2026 via E-Mail and FedEx. Merit ask consignees to take the following actions: 1. Ensure that applicable personnel within your organization are made aware of this safety notice. 2. If the product has been further distributed to other facilities, institutions, or manufacturers, please immediately share this safety notice in accordance with your company policies. Please consider communicating the following to your customers: a. For continued safe use of the Philips Bridge" and Philips Bridge" Plus Occlusion Balloons, as outlined in the Prophylactic Directions for Use in the IFU, the following actions should be taken prior to the beginning of any procedure to confirm proper interaction between devices: i. After placement of the guidewire, advance the Balloon over the guidewire. This will confirm that the Balloon can freely be advanced over the entire length of the guidewire without resistance, thereby confirming that the guidewire is compatible with the Balloon. ii. Should resistance be felt during this step, the guidewire should be replaced with an alternate 0.035 stiff/super stiff wire, and Step 1 should be repeated prior to the beginning of the procedure. 3. Please circulate this information to all users of the Philips Bridge", Philips Bridge" Plus, and Bridge" Prep Kit device, or to any organization where the affected product may have been transferred, so that they are aware of the issue. Merit encourages you to instruct your customers to post the safety notice on or near the affected products.
- Root cause
- Nonconforming Material/Component
- Status
- Open, Classified
- Product code
- OEQ
- Distribution
- US Nationwide distribution in the state of CO.
- Date initiated
- 2026-04-02
- Date posted
- 2026-04-29
- Location
- South Jordan, UT
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.