Philips Azurion Systems with software versions R2.1.10 and R2.2.10. Includes the following label descriptions and corres

FDA device recall Z-1091-2026 · posted 2026-01-14 · Open, Classified

Recall details

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Reason for recall
Six (6) identified software issues which may result in the following: system keeps restarting, AMC triple drive, C-Partition of Suite PC running out of free space, system remains in continuous restart mode after the start-up, misalignment of Market Tool Overlay, and longitudinal positional error.
Action taken
Philips notified consignees on about 12/12/2025 via certified mailing. Consignees were instructed to identify any affected systems on hand, circulate the noification to all users of the system for awareness, notify any recipients if the affected systems have been transferred to another organization, maintain the notification with documentation of the system until Philips completes the correction, and complete and return the response form provided. Philips also provided recommended actions for each issue, if encountered. Philips will address the identified issues by implementing a software update (R2.1.15 or R2.2.15) in all affected systems (FCO72200660 to 2.2.15, FCO72200672 to 2.1.15). Philips expects to have this software update released by Q4 2026 (subject to regulatory clearance). Local Philips representatives will contact each consignee to schedule a visit to install the software updates once available.
Root cause
Software design
Status
Open, Classified
Product code
OWB
Quantity affected
10,466 units
Distribution
Worldwide distribution - US Nationwide and the countries of Afghanistan, Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia an
Date initiated
2025-12-12
Date posted
2026-01-14
Location
Best, Netherlands

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Azurion R2.1 (K200917)Philips Medical Systems Nederland B.V.2020-05-01
Azurion R2.0 (K181830)Philips Medical Systems Nederland B.V.2018-08-02
Azurion series R1.2 (K172822)Philips Medical Systems Nederland B.V.2017-11-22