Philips Azurion Interventional Fluoroscopic X-ray Systems, specifically the Disk Bay of the PC component of the affected

FDA device recall Z-1152-2024 · posted 2024-02-16 · Open, Classified

Recall details

Recalling firm
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Reason for recall
Disk Bay may not perform as intended. System may stop functioning and imaging may not be possible, resulting in delay of procedure.
Action taken
On January 2, 2024, the firm notified affected consignees via" URGENT Medical Device Correction" letters. These letters covered issues with three components of the PCs that may not perform as intended and which may therefore lead to delay in treatment: 1) DIMMs, 2) Disk Bay, and 3) Framegrabber Card. This recall record is for the Disk Bay issue. Customers should circulate the notice to all users and establish an emergency protocol for if you experience the issue during a procedure. Customers were provided with monitoring activities to undertake while waiting for their system to be corrected. Philips will be replacing the three components in all affected systems. Philips will contact customers to schedule a visit to replace the affected components. If you need additional information or support concerning this issue, please contact your local Philips representative at 1-800-722-9377 (Philips Customer Care Solutions Center).
Root cause
Process design
Status
Open, Classified
Product code
OWB
Quantity affected
13, 133 (3,408 US; 9,725 ROW)
Distribution
Worldwide Distribution: US nationwide including states of: AL, AK, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NH, NJ, NM, NV, NY, OH, O
Date initiated
2024-01-02
Date posted
2024-02-16
Location
Best, Netherlands