Atlas FDA

Pharmaseal Thoracentesis Tray with Catheter, catalog number 4341B; a sterile, single use, Rx, latex free procedure tray

FDA device recall Z-0420-03 · posted 2003-01-09 · Terminated

Recall details

Recalling firm
Cardinal Health
Reason for recall
Raw material used in the manufacture of catheters may contain a contaminant predisposing the catheter to become brittle, possibly breaking during use.
Action taken
Recall letters dated 11/15/02 were sent via 2-day FedEx to the direct accounts. The accounts were informed of the potential for breakage of the catheters during use, requested to inspect their inventories for the affected lot numbers, and segregate any found for return to Allegiance for replacement. Distributors were requested to sub-recall the product from their accounts. The recall was extended to include the 21 additional lots by letter dated 4/8/04. The letters were sent via Airborne Express, 1-Day, on 4/8/04. The instructions were the same as the 11/15/02 letter, except that Allegiance has changed their name to Cardinal Health.
Root cause
Other
Status
Terminated
Product code
GCB
Quantity affected
232,228 trays
Distribution
Nationwide distribution, including Puerto Rico and Guam, and internationally to Saudi Arabia
Date initiated
2002-11-15
Date posted
2003-01-09
Date terminated
2006-04-14
Location
McGaw Park, IL