Pharmaseal Thoracentesis Tray with Catheter, catalog number 4341B; a sterile, single use, Rx, latex free procedure tray
FDA device recall Z-0420-03 · posted 2003-01-09 · Terminated
Recall details
- Recalling firm
- Cardinal Health
- Reason for recall
- Raw material used in the manufacture of catheters may contain a contaminant predisposing the catheter to become brittle, possibly breaking during use.
- Action taken
- Recall letters dated 11/15/02 were sent via 2-day FedEx to the direct accounts. The accounts were informed of the potential for breakage of the catheters during use, requested to inspect their inventories for the affected lot numbers, and segregate any found for return to Allegiance for replacement. Distributors were requested to sub-recall the product from their accounts. The recall was extended to include the 21 additional lots by letter dated 4/8/04. The letters were sent via Airborne Express, 1-Day, on 4/8/04. The instructions were the same as the 11/15/02 letter, except that Allegiance has changed their name to Cardinal Health.
- Root cause
- Other
- Status
- Terminated
- Product code
- GCB
- Quantity affected
- 232,228 trays
- Distribution
- Nationwide distribution, including Puerto Rico and Guam, and internationally to Saudi Arabia
- Date initiated
- 2002-11-15
- Date posted
- 2003-01-09
- Date terminated
- 2006-04-14
- Location
- McGaw Park, IL