Persona The Personalized Knee System 2.5 mm Female Screw 48 mm Length Sterile Qty-2. Sterilized using irradiation, singl
FDA device recall Z-2737-2014 · posted 2014-09-30 · Terminated
Recall details
- Recalling firm
- Zimmer, Inc.
- Reason for recall
- The Persona 48mm X 2.5mm Female Screw is being recalled due to potential use of the screw in cortical bone, and corresponding stripping of the 2.5mm Persona Hex Driver.
- Action taken
- Zimmer sent an URGENT MEDICAL DEVICE RECALL notifications dated September 9, 2014, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. 1. Review the notification and the updated portion of the Persona surgical technique provided in Attachment 1, and ensure affected personnel are aware of the contents. 2. Locate and quarantine any affected product listed above and notify your Zimmer sales representative. 3. Your Zimmer sales representative will remove the recalled product from your facility. 4. If after reviewing this notification you have further questions or concerns please call the customer call center at 1-877-946-2761 between 8:00 am and 5:00pm EST.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- GFC
- Quantity affected
- 16,380 units
- Distribution
- Nationwide Distribution including the states of : HI, CA, UT, NY, OH, IN, AL, FL, GA, WI, IA, MN, NE, SD, NJ, PA, ND, IL, MO, VA, MD, NC, MI, OR, WA, KS, KY, TN, WV, OK, CT, MA, ME, AZ, NM, TX, LA, ID
- Date initiated
- 2014-09-09
- Date posted
- 2014-09-30
- Date terminated
- 2015-05-05
- Location
- Warsaw, IN
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.