PERPOS PLS, CATALOG #'s 9045-01 & 9045-02 A sterile, medical device kit consisting of the components listed in Table 1-1

FDA device recall Z-3033-2011 · posted 2011-08-18 · Terminated

Recall details

Recalling firm
Interventional Spine Inc
Reason for recall
This recall has been initiated due to Pull Pins that may disengage from the Facet Screw during the compression step, requiring that compression be completed with a device driver rather than the compression tool. Use of this product may result in less than optimal compression leading to pseudoarthrosis, non-union or facet pain.
Action taken
Interventional Spine, Inc. sent a recall letter dated March 15, 2011, informing consignees of the product, the problem, and the action to be taken. Consignees were asked to immediately examine their inventory and quarantine product subject to recall. In addition, if they distributed the product further they were asked to notify Jane Metcalf, the Vice President of Quality Assurance at once. Interventional Spine further informed consignees that the recall was being carried out to the individual consignee level. And their assistance was appreciated and necessary to prevent use of this product. They were then asked to complete and return the enclosed response form as soon as possible. Questions were directed to call their regional directors. For questions regarding this recall call 949-472-0006.
Root cause
Other
Status
Terminated
Product code
MRW
Quantity affected
553 units
Distribution
Nationwide Distribution including CO, CT, FL, KY, LA, MA, MO, NC, NY, OK, OR, TN, TX, VA & WI
Date initiated
2011-03-15
Date posted
2011-08-18
Date terminated
2013-02-11
Location
Irvine, CA