Perfusion System 8000, base 5 pump, 220V/240V safety monitor included The Sarns Modular Perfusion System 8000 is indicat
FDA device recall Z-0129-2013 · posted 2012-10-24 · Terminated
Recall details
- Recalling firm
- Terumo Cardiovascular Systems Corporation
- Reason for recall
- Terumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2007 of various malfunctions of the power distribution and battery back-up systems for the Sarns Modular Perfusion System 8000. Very few of the reports occurred during cardiopulmonary bypass; the majority of the malfunctions occurred during set-up or were reported during routine preventative maintenance. The malfunction
- Action taken
- TERUMO sent an URGENT MEDICAL DEVICE letter dated September 25, 2012, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed review the Medical Device Recall notice, assure that all users are aware of the notice and place the addendum in the Operator's Manual before page 1.13. Customers were also instructed to confirm receipt of the Urgent Medical Device letter by faxing, the attached Customer Response Form to 1-800-292-6551. For questions customers were instructed to call 1-800-521-2818.
- Root cause
- Labeling design
- Status
- Terminated
- Product code
- DTQ
- Quantity affected
- 312 units
- Distribution
- Worldwide Distribution - USA including AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NY, OH, OK, OR, PA, SC, TN, TX, VA, WA, WI, W
- Date initiated
- 2012-09-25
- Date posted
- 2012-10-24
- Date terminated
- 2015-08-21
- Location
- Ann Arbor, MI
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| SARNS 8000 MODULAR PERFUSION SYSTEM (K915183) | 3M Health Care, Sarns | 1992-07-08 |