Atlas FDA

Perfusion System 8000, base 4 pump, 100V safety monitor included The Sarns" Modular Perfusion System 8000 is indicated f

FDA device recall Z-0284-2013 · posted 2012-11-13 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corporation
Reason for recall
Terumo Cardiovascular Systems (Terumo CVS) has received multiple reports since 2007 of various malfunctions of the arterial and cardioplegia monitors for the Sarns Modular Perfusion System 8000. The malfunctions can be broadly categorized as follows monitor power malfunctions, inability to measure pressure and inaccurate and/or false pressure alarms. If the malfunction of the arterial or cardio
Action taken
The firm, Terumo CVS, sent an "URGENT MEDICAL DEVICE RECALL" notice/Safety Advisory Alert dated September 24, 2012 and an addendum to the Operator's Manual to alert all affected Customers. The notice described the product, problem and actions to be taken. The customers were instructed to review this medical device recall notice; assure that all users are aware to this notice; place the addendum in the Operator's Manual just before page 4.1 and complete and return the attached Customer Response Form via Fax to: 1-734-741-6149 or email to:www.termuo-cvs.com. If you have any questions or concerns contact Terumo CVS Customer Service at 1-800-521-1818; Monday - Friday, 8 AM - 6 PM ET.
Root cause
Device Design
Status
Terminated
Product code
DTQ
Quantity affected
26
Distribution
Worldwide distribution: USA (nationwide) including states of: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NH, NJ, NM, NY, OH, OK, OR, PA
Date initiated
2012-09-25
Date posted
2012-11-13
Date terminated
2013-07-11
Location
Ann Arbor, MI

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
SARNS 8000 MODULAR PERFUSION SYSTEM (K915183)3M Health Care, Sarns1992-07-08