Pentax Medical Video Processor; Model Number: EPK-i8020c;
FDA device recall Z-2329-2025 · posted 2025-08-15 · Open, Classified
Recall details
- Recalling firm
- Pentax of America Inc
- Reason for recall
- During endoscopic procedures using a combination of the video processor EPK-i8020c and i20c series video endoscope, the observed image can become reddish or dark. Users have observed smoke like steam and noted that the light guide at the tip is hot during use / after the endoscope is removed from the patient and may cause thermal injury to the patient s mucous membrane.
- Action taken
- On 7/16/2025, "URGENT MEDICAL DEVICE CORRECTION" letters were sent to customers. Customer Instructions: PENTAX Medical will conduct the software update on the affected devices at your facility. Your local PENTAX Medical representative will contact you to schedule the required updates for your equipment. " Please download the Instructions for Use (IFU) for the EPK-i8020c Video Processor from the PENTAX Medical online IFU library at https://ifu.pentaxmedical.com. " Please complete the enclosed Field Correction Response Form upon receipt of this package, and email to PENTAX Medical at customeradvisories@pentaxmedical.com. Contact Information: Please indicate through the attached response form that you have received and understood this information, by completing it and returning it no later than July18, 2025 at: customeradvisories@pentaxmedical.com If you have any questions regarding this action, please feel free to contact us at: " Tel: 1-800-431-5880 (8:30 AM 5:00 PM, Monday Friday, EST) " Fax: (800)-579-5432) " Email: customeradvisories@pentaxmedical.com
- Root cause
- Device Design
- Status
- Open, Classified
- Product code
- PEA
- Quantity affected
- 144 units
- Distribution
- US Distribution to states of: AZ, CA, DC, FL IA, ID, IL, KS, MA, MD, MI, MO, NC, NJ, NM, NY, OK, PA, TX, WA.
- Date initiated
- 2025-07-16
- Date posted
- 2025-08-15
- Location
- Montvale, NJ
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| PENTAX Medical Video Processor (EPK-i8020c) (K251127) | Pentax of America, Inc. | 2025-06-03 |
| PENTAX Medical Video Processor EPK-i8020c, PENTAX Medical Video Upper GI Scope EG29-i20c, PENTAX Medical Video Colonoscope EC38-i20cL (K231249) | Pentax of America, Inc. | 2023-07-28 |