Penner Manufacturing Transfer Electric Bariatric with Scale, Patient Transfer/Lift System, Cascade Transfer, Models 3835
FDA device recall Z-1915-2011 · posted 2011-04-08 · Terminated
Recall details
- Recalling firm
- Penner Mfg Inc
- Reason for recall
- Incomplete weld on the pillar mounting bracket
- Action taken
- The firm, Penner, sent an undated letter flagged "Important Safety Inspection: Action Required" via certified mail on January 19, 2009 to its customers. The letter described the product, problem and actions to be taken. The customers were instructed to inspect the Penner Patient Transfer for any missing Welds of the Pillar Mounting Bracket; inspect the areas indicated in the enclosed photos to confirm the bracket is welded on all edges to the main frame of the base; complete the inspection, and complete and return the attached WELD INSPECTION RESULTS OF PENNER PATIENT TRANSFER form via fax to 402 694-5844 or mail to Penner Manufacturing Inc, 102 Grant Street, Aurora, NE 68818, Attn: Quality Assurance Manager. Penner stated in the letter, if welds are not complete (no weld), discontinue use and contact the Quality Assurance Manager at Penner Manufacturing immediately at 402-694-5003. If you have any questions, contact the Quality Assurance Manager at 402-694-5003.
- Root cause
- Employee error
- Status
- Terminated
- Product code
- FNG
- Quantity affected
- 20 lifts
- Distribution
- Worldwide distribution: USA and countries including: Canada and Mexico.
- Date initiated
- 2009-01-19
- Date posted
- 2011-04-08
- Date terminated
- 2011-04-08
- Location
- Aurora, NE