Pelton & Crane Spirit 3000 dental chair (Model #SP30)
FDA device recall Z-1850-2011 · posted 2011-03-30 · Terminated
Recall details
- Recalling firm
- Pelton & Crane Company
- Reason for recall
- The Link Arm Assembly that connects some rear mounted options to the Spirit 3000 or Spirit 1800 dental chairs may potentially break and fall off the dental chair.
- Action taken
- Pelton & Crane sent an "Urgent Medical Device Recall" letter dated 3/04/2011. They were instructed to: Locate the affected devices this recall is in regards to. Schedule service with the end user for Pelton & Crane representative to replace the affected components this recall is in regards to. International accounts were instructed to order link arm assemblies from Pelton & Crane and replace the affected components. All consignees were to complete the Recall Acknowledgement/Return Form and return the form to Pelton & Crane.
- Root cause
- Process control
- Status
- Terminated
- Product code
- NRU
- Quantity affected
- 9 units
- Distribution
- Distribution to USA, Canada, and the Netherlands.
- Date initiated
- 2011-03-04
- Date posted
- 2011-03-30
- Date terminated
- 2011-08-16
- Location
- Charlotte, NC