Atlas FDA

Pediatric Craniotome Ref:CRANI-P-G1

FDA device recall Z-0924-2023 · posted 2023-01-07 · Open, Classified

Recall details

Recalling firm
The Anspach Effort, Inc.
Reason for recall
Ball bearings in the CRANI-A (ANSPACH Power Tools Adult Craniotome Attachment) came out of the attachment, possibly during removal of the attachment, intra-operatively. Failure to follow the recommended inspection intervals provided in the IFU may result in serious patient injury
Action taken
DePuy Synthes initiated "URGENT: MEDICAL DEVICE CORRECTION (Notification)" via UPS on 12/07/22. Letter states reason for recall, health risk an action to be taken. The customers were instructed to do the following: Review your inventory and make a plan to follow the suggested manufacturer inspection interval of 12 months associated with your product, as specified in the IFU. " Forward this notice to anyone in your facility that needs to be informed (i.e., those who manage, transport, store, stock, or use the subject products). " If any of the subject products have been forwarded to another facility, contact that facility and provide them with this notice. " Once this notification is received, please complete, sign, and return the attached Business Response Form (page 3 of this letter) via Fax: 877-496-5042 or Scan/email: Synthes3606@sedgwick.com within 3 business days of receipt of this notification. Contact your Product Support Team through email at ra-dpyus-prodsupport@its.jnj.com or via phone at 1-800.327.6887, option #2 for additional support as needed.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
HBC
Quantity affected
6059 units in total
Distribution
US Nationwide and OUS Foreign: Impacted Foreign Countries Australia Bangladesh China Hongkong India Indonesia Japan Korea (Republic of) Malaysia New Zealand Philippines Singapore Taiwan, Province of C
Date initiated
2022-12-07
Date posted
2023-01-07
Location
Palm Beach Gardens, FL

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ANSPACH EG1 HIGH SPEED SYSTEM, G1 ATTACHMENTS (K133604)The Anspach Effort, Inc.2014-01-10
ANSPACH EMAX DRILL SYSTEM (K011444)The Anspach Effort, Inc.2001-08-08