PCF-H190DL, EVIS EXERA III COLONOVIDEOSCOPE
FDA device recall Z-0418-2024 · posted 2023-11-29 · Open, Classified
Recall details
- Recalling firm
- Aizu Olympus Co., Ltd.
- Reason for recall
- Specific devices may have an improperly repaired light guide connector due to missing adhesive to secure the suction channel mouthpiece to the scope connector. Issue may result in loss of image, decreased suction, or impact the ability to effectively reprocess the device. This in turn may lead to unanticipated delay to treatment, or patient exposure to contaminated material and subsequent infection.
- Action taken
- On October 16, 2023, the firm notified affected customers through URGENT MEDICAL DEVICE REMOVAL letters. Customers were instructed to inspect their inventory and identify affected devices. Affected devices should be quarantined and not used. Olympus Customer Solutions will contact customers by phone and email to arrange the return and repair of the affected product. If you require additional information, please do not hesitate to contact Olympus at 647-999-3203 or Cynthia.Ow@Olympus.com.
- Root cause
- Process control
- Status
- Open, Classified
- Product code
- FDF
- Quantity affected
- 24 incorrectly repaired units
- Distribution
- Domestic distribution to FL, KY, MA, NJ, NY, and PA
- Date initiated
- 2023-10-16
- Date posted
- 2023-11-29
- Location
- Aizuwakamatsu, 3 Chome, Japan