Patient Side Cart assembly, 4-arm, for use with Intuitive Surgical, Inc. da Vinci Surgical System IS 4000. The patient-s

FDA device recall Z-1718-2014 · posted 2014-06-04 · Terminated

Recall details

Recalling firm
Intuitive Surgical, Inc.
Reason for recall
Torque wrenches used to tighten certain PSC bolts were out of calibration. This could lead to over- or under-tightening of bolts that may result in system faults, mechanical sound (e.g. scraping, squeaking, rattling), cracks or vibration in the system.
Action taken
Intuitive sent an Urgent Medical Device Correction letter dated May 13, 2014 to affected customers. The letter identified the affected product, problem and actions to be taken. All personnel who use the device should be notified of the issue and the acknowledgement form completed and returned. Intuitive reps will visit each affect sites to install a replacement PSC and ensure that it is properly integrated with the Da Vinci. For questions contact Intuitive Surgical at 800-876-1310 Option 3 (Monday - Friday 6 am to 5 pm PST).
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
NAY
Quantity affected
2
Distribution
Medical facilities in TX and FL
Date initiated
2014-05-09
Date posted
2014-06-04
Date terminated
2014-06-04
Location
Sunnyvale, CA