PASV Central Venous Catheter, 7F, Cat. # CVC702IK, Model # M0001CVC702IK. The 7Fr PASV Central Venous Catheter is a pack

FDA device recall Z-1112-05 · posted 2005-08-10 · Terminated

Recall details

Recalling firm
Boston Scientific Corporation
Reason for recall
Kits labeled to contain 0.028" guidewire contain 0.038" guidewire.
Action taken
Letters dated May 25, 2004 with instructions to return product to the Quincy, MA location.
Root cause
Other
Status
Terminated
Product code
LJS
Quantity affected
290
Distribution
278 units shipped to 31 consignees throughout the U.S. 12 units shipped to one foreign consignees located in the UK.
Date initiated
2004-05-25
Date posted
2005-08-10
Date terminated
2006-02-06
Location
Natick, MA