Para-Pak Zn-PVA/10% Formalin-for the routine collection, transportation, preservation, and examination of stool specimen
FDA device recall Z-0520-2024 · posted 2023-12-08 · Open, Classified
Recall details
- Recalling firm
- Meridian Bioscience Inc
- Reason for recall
- Plastic vials contain a manufacturing defect may cause an increased risk of leakage and presents a risk of incidental exposure to biological hazards associated with direct contact with the patient specimen.
- Action taken
- Meridan issued Urgent Medical Device Correction letters via email on 10/19/23 to Distributors and End-Users. Letter states reason for recall, health risk and action to take: Identify and segregate any remaining inventory. REQUIRED ACTIONS FOR PARA-PAK CLEAN VIAL, CATALOG 900312, LOT 509940Q: " Identify and segregate any remaining inventory. " Discard remaining Inventory of Para-Pak Clean Vial, Catalog 900312, Lot 509940Q in accordance with local, state, and/or federal regulations. Record quantity discarded. o Contact Meridian Technical Support at 1-800-343-3858 to coordinate product replacement activities. " Complete and return the Customer Notification Form. This will indicate notification of this field correction. Complete this form even if you have no remaining inventory. REQUIRED ACTIONS FOR PARA-PAK Zn-PVA/10% FORMALIN, CATALOG 301012, LOT 509130Q: " Identify and segregate any remaining inventory. " For remaining inventory of Para-Pak Zn-PVA/10% Formalin, Catalog 301012, Lot 509130Q, inspect each individual vial set for signs of leakage. o Leakage would be able to be visually observed through inspection of the Ziplock-style bag containing the two vials of preservatives. Signs of leakage include identification of preservative outside of the plastic vials or at the bottom of the Ziplock-style bag. DO NOT remove vials that exhibit leaking from the Ziplock-style bag. Handle with care to prevent direct contact of the preservative with the skin or eyes. o Record the quantity of 2 vial sets that demonstrated signs of leakage. " Discard the remaining inventory of Para-Pak Zn-PVA/10% Formalin, Catalog 301012, Lot 509130Q in accordance with local, state, and/or federal regulations. Record quantity discarded. o Contact Meridian Technical Support 1-800-343-3858 to coordinate product replacement activities. " Complete and return the Customer Notification Form. This will indicate notification of this field correction. Complete this form even if you have no remaining invento
- Root cause
- Process control
- Status
- Open, Classified
- Product code
- IGG
- Quantity affected
- 201 Cases
- Distribution
- Nationwide Foreign: Italy
- Date initiated
- 2023-10-19
- Date posted
- 2023-12-08
- Location
- Cincinnati, OH