Atlas FDA

Panoramic X-ray Model PC-1000 The PC-1000 will enable the user to take panoramic x-ray images. The PC-1000/Laser 1000 wi

FDA device recall Z-2538-2015 · posted 2015-09-01 · Terminated

Recall details

Recalling firm
Panoramic Rental Corp.
Reason for recall
Jackscrew nut part failure potentially leading to device malfunction and patient/operator injury.
Action taken
The firm, Panoramic Corporation, sent an "URGENT: MEDICAL DEVICE RECALL" letter dated 7/17/2015 via certified mail to customers on 7/20/2015, with response forms for customers to complete and return. The firm previously issued "IMPORTANT SERVICE BULLETIN" notices in 2003 and 2008 via FED-EX. The letter describes the product, problem and actions to be taken. The customers are instructed to select to use several options including replacing the failing part with one containing the safety feature, or trading in their machine for a newer one; discontinue using the devices immediately regardless the option you select, and complete and return the attached response card via email to recall@pancorp.com or fax to 260-489-5683. If the machine has been sold or donated to another dental outfit, the customer should notify Panoramic so that they can update their records. When the customer is reached, the firm will confirm whether or not the machine is still in service. Customers with questions may contact 800-654-2027 Monday through Thursday 8:00 AM to 6:00 PM and Friday 8:00 AM to 5:00 PM EST to discuss these options and determine which is best for their practice.
Root cause
Equipment maintenance
Status
Terminated
Product code
EHD
Quantity affected
7885
Distribution
Worldwide distribution: US (nationwide) all 50 states, including Puerto Rico; and countries of Antigua, Armenia, Aruba, Austria, Canada, China, Colombia, Ecuador, Egypt, England, France, Germany, Gree
Date initiated
2015-06-04
Date posted
2015-09-01
Date terminated
2017-05-30
Location
Fort Wayne, IN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
LASER - 1000 (K882436)Panoramic Corp.1988-08-04
PANORAMIC X-RAY MODEL PC-1000 (K870236)Panoramic Corp.1987-02-13