P21x/5-1 MHz transducer, Part Number P07698-23 Product Usage: The P21x/5-1 MHz phased array transducer is designed for c

FDA device recall Z-0027-2014 · posted 2013-10-18 · Terminated

Recall details

Recalling firm
SonoSite, Inc.
Reason for recall
FUJIFILM SonoSite Inc. remove P21x/5-1 Transducer because the firm discovered the Biopsy Bracket Sensor is not working as intended.
Action taken
SonoSite started calling their consignees on September 13, 2013. For customers who could not be reached via phone, a letter Urgent -Medical Device Removal will be sent on September 17, 2013. Affected customers will be offered a replacement transducer (P07698-24) free of charge. Consignees in the U.S. can call 1-877-657-8118, Monday-Friday, 6:00 a.m. - 5:00 p.m. (PST) or send E-mail to service@sonosite.com. Customers outside the US can call 425-951-1200.
Root cause
Error in labeling
Status
Terminated
Product code
IYO
Quantity affected
10 units
Distribution
Worldwide Distribution - USA Nationwide in the states of Florida, Georgia, Oklahoma, Pennsylvania, Tennessee, Texas, West Virginia, and in the country Japan and Italy.
Date initiated
2013-09-11
Date posted
2013-10-18
Date terminated
2013-12-11
Location
Bothell, WA