OxySure Model 615 disposable replacement cartridges, for portable oxygen generator.

FDA device recall Z-2253-2015 · posted 2015-07-24 · Terminated

Recall details

Recalling firm
Oxysure Systems, Inc.
Reason for recall
Flow rate of the recalled replacement cartridges does not meet the draft FDA standard of a minimum 6 LPM for a minimum of 15 minutes.
Action taken
The recalling firm will notify their affected customers via email beginning the week of 6/15/15. The recall notification will have a response for to be completed by the consignee. The recall notification calls for the customers to dispose of the affected product according to the MSDS.
Root cause
Under Investigation by firm
Status
Terminated
Product code
CAW
Quantity affected
1005 units
Distribution
Nationwide throughout the US and Chile, Trinidad, China
Date initiated
2015-06-12
Date posted
2015-07-24
Date terminated
2020-10-19
Location
Frisco, TX