Oxygen Masks- Labeling as follows: MADA, Inc. Adult Medium Concentration Oxygen Mask with Non-crush 7 Coiled Tubing, Con
FDA device recall Z-1256-2011 · posted 2011-02-15 · Terminated
Recall details
- Recalling firm
- Convatec Inc.
- Reason for recall
- Face tents and Oxygen masks may have a sticky glue-like substance on the inside or outside of the mask consisting of dioctyl phthalate, epoxidized soybean oil and glycerol monoleate.
- Action taken
- Unomedical, Inc issued recall letters with attached return response forms dated January 7, 2011 via UPS. The letters request the customer stop use and quarantine product. The recall response form is to be completed and returned to Unomedical. The letters/return response form were also sent out in Spanish. Another recall letter was sent dated February 16, 2011 to capture additional lots that werent included in the initial letter.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- BYG
- Quantity affected
- 1113200 pcs.
- Distribution
- Worldwide distribution, including USA, Puerto Rico, Canada, Mexico, Colombia, El Salvador, Paraguay, UK, Austria, Belgium, Cyprus, Czech Republic, Denmark, Estonia, Faroe Islands, Finland, France, Ger
- Date initiated
- 2011-01-07
- Date posted
- 2011-02-15
- Date terminated
- 2013-02-27
- Location
- Skillman, NJ