Atlas FDA

Oxoid dry Spot Staphytect Plus, Catalog #DR0100M, 120 tests/kit. Oxoid Ltd., Wade road, Basingstoke, Hants, UK. Latex sl

FDA device recall Z-0817-2010 · posted 2010-02-23 · Terminated

Recall details

Recalling firm
Remel, Inc
Reason for recall
The products may give false positive results with some non-Staphylococcus aureus strains
Action taken
Thermo Fischer Scientific, Remel Inc. issued letters dated July 20, 2009 via regular mail explaining the reason for recall, risks involved, and requested they discard all remaining product. A Product Inventory Checklist was enclosed to report the amount of product their customer discarded. A second letter dated October 19, 2009 was issued via certified mail. For further information, contact Thermo Fischer Scientific, Remel Inc. at 1-800-255-6730, Extension 3304185.
Root cause
Process control
Status
Terminated
Product code
JWX
Quantity affected
27 kits
Distribution
Nationwide Distribution -- MN, CT, FL, GA, CA, NY, MI, MO, PA, and IL.
Date initiated
2009-07-20
Date posted
2010-02-23
Date terminated
2012-04-25
Location
Lenexa, KS