Oxoid Antimicrobial Susceptibility Test Discs, CTO425B Ciprofloxacin CIP 5 mcg, 50 discs per cartridge, 5 cartridges per

FDA device recall Z-0528-2014 · posted 2013-12-17 · Terminated

Recall details

Recalling firm
Remel Inc
Reason for recall
Boxes labeled as CIP5 may contain cartridges of AMP5.
Action taken
Thermo Fisher sent a Field Safety Notice dated November 13, 2012, to all affected customers. The letter identified the product, the problem and the action to be taken by the customer. Customers were instructed to check stocks of the affected product if necessary and contact Customer Service or their local Oxoid supplier who will issue a replacement. The notice should be passed on to all who need to be aware within their organization. Customers were also instructed to discard any product labeled as AMP5. For questions, customers were instructed to call +44 (0)1256 841144.discard any product labeled AMP5. For questions regarding this recall call 913-895-4185.
Root cause
Mixed-up of materials/components
Status
Terminated
Product code
JTN
Quantity affected
60/5/50-disc cartridge boxes
Distribution
Nationwide Distribution including AL, CA, CO, FL, IA, KS, LA, MA, MS, NE, NH, NJ, NY, OK, OR, RI, TN, TX, VA, WA WI, and Puerto Rico.
Date initiated
2012-11-21
Date posted
2013-12-17
Date terminated
2014-03-12
Location
Lenexa, KS