Oxford Fixed Lateral Bearing Size F Right, Part Number 154375 Lot 744260, Sterile, The product is placed into an Aluminu
FDA device recall Z-0956-2017 · posted 2017-01-06 · Terminated
Recall details
- Recalling firm
- Zimmer Biomet, Inc.
- Reason for recall
- Zimmer Biomet is initiating a removal of a single lot of Oxford Fixed Lateral Bearings due to mislabeling. Zimmer Biomet received product complaints reporting that when the package labeled as a Right was opened it actually contained a Left Bearing.
- Action taken
- Zimmer Biomet sent an Urgent Medical Device Recall letter dated December 6, 2016. to all affected customers for a single lot of Oxford Fixed Lateral Bearings via e-mailed letter. Zimmer Biomet received two product complaints reporting that when the package labeled as a Right was opened it actually contained a Left Bearing. Risks: If another bearing is not available, a delay greater than 30 minutes may occur. If the product of the incorrect orientation is implanted, premature revision may be required. Customers were instructed to review the recall notification and ensure affected team members are aware of the contents. 2. Immediately locate and quarantine affected product in your inventory. 3. Complete the Certification of Acknowledgement portion of the Inventory Return Certification Form a. Return a digital copy within three (3) days via email to corporatequality.postmarket@zimmerbiomet.com. 4. Immediately return all affected product from your distributorship and affected hospitals within your territory along with a completed Inventory Return Certification Form to Zimmer Biomet. Request a Return Authorization Number via email at rgarequest@zimmerbiomet.com . Be sure to specify RECALL as the RGA type. b. For each return, send a copy of Inventory Return Certifictaion Form via email to corporatequality.postmarket@zimmerbiomet.com. c. Include a hardcopy of Inventory Return Certifictaion Form to Zimmer Biomet with your shipment for immediate processing. d. Mark the outside of the returns box(es) clearly with RECALL. 5. If after reviewing this notice you have further questions or concerns please call 1-877-946-2761 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of business hours will receive a prompt to record a voicemail and will be returned upon receipt. Alternatively, your questions may be sent by email to corporatequality.postmarket@zimmerbiomet.com. Customer Service Zimmer Biomet, Inc. P.O. Box 708 Warsaw, I
- Root cause
- Employee error
- Status
- Terminated
- Product code
- HRY
- Quantity affected
- 12 units distributed
- Distribution
- Domestic: None Froeign: Netherlands & Switzerland VA/DOD: None
- Date initiated
- 2016-12-06
- Date posted
- 2017-01-06
- Date terminated
- 2017-06-01
- Location
- Warsaw, IN