Outlook ES pump is intended for use with B. Braun Medical Inc. Horizon Pump IV Sets. Used to regulate the flow of primar

FDA device recall Z-2164-2013 · posted 2013-09-06 · Terminated

Recall details

Recalling firm
B Braun Medical, Inc.
Reason for recall
This notice is being provided in follow up to the field correction initiated by B. Braun Medical Inc. on 26 August, 2011 due to an issue in the Outlook ES Safety Infusion System where the Care Area can inadvertently be exited thereby removing DoseGuard dose limit protection. This occurs when a sequence of HOLD - HOLD key strokes are performed on the pump key panel while the user is in the Dose Dat
Action taken
The firm decided to conduct a field correction to correct the issue and carried it out through routine field services. The firm was notified by FDA that this is a reportable event, so they notified their consignees via letters dated 10/31/2012.
Root cause
Software design
Status
Terminated
Product code
FRN
Quantity affected
1,970 pumps added for this expansion
Distribution
Nationwide Distribution.
Date initiated
2012-10-31
Date posted
2013-09-06
Date terminated
2016-02-03
Location
Carrollton, TX