Outlook 400ES Safety Infusion System, Model Number 621-400ES. For use with B. Braun Medical Inc. Horizon Pump IV Sets

FDA device recall Z-0837-2011 · posted 2011-01-10 · Terminated

Recall details

Recalling firm
B Braun Medical, Inc
Reason for recall
Possible pump stop with audible alarm, but no visual error indication that the pump is not delivering the medication. Under certain conditions in a network environment, Management Processor may become non-responsive, ceasing normal operation.
Action taken
B. Braun Medical Inc. notified consignee by letter on August 24, 2010.
Root cause
Component design/selection
Status
Terminated
Product code
FRN
Quantity affected
2,107 Pumps (582 Pumps to Winchester)
Distribution
Virginia and New York
Date initiated
2010-08-24
Date posted
2011-01-10
Date terminated
2014-05-29
Location
Carrollton, TX