Atlas FDA

OSTEORAPTOR CURVED 2.3 mm Suture Anchor SA ULTRABRAID COBRAID Suture, Blue; Part Number: 72203290 Biodegradable suture a

FDA device recall Z-2228-2013 · posted 2013-09-17 · Terminated

Recall details

Recalling firm
Smith & Nephew, Inc. Endoscopy Division
Reason for recall
Pinholes detected in a small number of primary pouches that may constitute a breach of the sterile barrier.
Action taken
Smith&Nephew sent an Urgent Product Recall Notification letter dated August 6, 2013, via Federal Express to all affected customers. Addendum customer letters for 4 additional part numbers were sent via Federal Express on August 19, 2013, The letters identified the product, the problem, and the action to be taken by the customer. Customers were asked to inspect their inventory, locate unused devices and quarantine them. Customers were also asked to complete the last three columns in the inventory Return Certification form indicating the quantities that need to be returned, and include their contact information in the spaces provided. Customers were instructed to contact Smith & Nephew's Demo Service Departmentat 800-343-5717 for a Return Authorization (RA#), and provide the information response form. Replacement product would be provided. For questions regarding this recall call 978-749-1073.
Root cause
Packaging
Status
Terminated
Product code
JDR
Quantity affected
10296 US
Distribution
Worldwide Distribution - USA (nationwide) Canada, Europe: Austria, Belgium, Cyprus, Czech Republic, Denmark, Germany, Greece, Finland, France, Italy, Ireland, Luxembourg, Netherlands, Norway, Poland,
Date initiated
2013-08-06
Date posted
2013-09-17
Date terminated
2016-04-20
Location
Andover, MA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
OSTEORAPTOR SUTURE ANCHOR (K082215)Smith & Nephew Inc., Endoscopy Division2008-11-03