Atlas FDA

Ortho Clinical Diagnostics VITROS Chemistry Products Vapor Adsorption Cartridge REF/CAT No. 680 0100. VITROS Vapor Adsor

FDA device recall Z-0755-2013 · posted 2013-01-30 · Terminated

Recall details

Recalling firm
Ortho-Clinical Diagnostics
Reason for recall
Some VITROS Chemistry Products Vapor Adsorption Cartridge CAT No. 6800100 cartons incorrectly contain a Particulate Cartridge Part No. J19612. In some instances, the Particulate Cartridge has been inadvertently installed on the VITROS System instead of the intended VITROS Vapor Adsorption Cartridge which may bias results.
Action taken
Ortho-Clinical Diagnostics sent URGENT PRODUCT CORRECTION Letters (dated 12/13/12) along with Q&A and Confirmation of Receipt Forms to distributors and consignees on 12/13/12. They were advised of the issue and instructed to discard their current inventory of VITROS Vapor Adsorption Cartridges. On 12/13/12, foreign affiliates were informed by e-mail of the issue and instructed to notify their consignees of the issue and actions.
Root cause
Packaging process control
Status
Terminated
Product code
JJE
Quantity affected
Domestic: 15,891; Foreign: 5237
Distribution
Worldwide Distribution, including Nationwide (USA) and the countries of AU, GM, FR, IT, SP, BR, BM, AR, CA, JA, MX, UK, CH, CL, CO, IN, EC, PE, PY, PA, PT, NZ, SG, VE and UY.
Date initiated
2012-12-13
Date posted
2013-01-30
Date terminated
2018-07-05
Location
Rochester, NY

Monitor recalls programmatically