ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in Radiation
FDA device recall Z-0398-2024 · posted 2023-11-21 · Open, Classified
Recall details
- Recalling firm
- LEONI CIA CABLE SYSTEMS
- Reason for recall
- Users of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when the movements are again allowed after an unforeseen interruption signal of motion may result in a collision of the patient with another stationary component of the treatment room.
- Action taken
- BizLink s notified customer via email with delivery and read confirmation and with the Field Safety Notice (FSCA Ref: UMCU-638) on October 6, 2023. Letter states reason for recall, health risk and action to take: BizLink technical support team will contact customers shortly to organize the deployment of the hot fix in the proton-therapy treatment centers. While waiting for the hot fix, BizLink recommend that customers implement the following action at the customer software level to avoid this problem: Send a stopFast command just after reactivating the motionEnable / the E-stop / the collision detection by Safety Laser and before sending any other motion command. A hot fix to correct this bug has been developed. BizLink shall provide this hot fix (version OSS 6.1 and version OSS 7.3) and the detailed documentation so that customers can deploy the new software version on clinical sites available on October 27, 2023 at the latest. If you require further clarification, please feel free to contact our BizLink Team (bfrc_CustomerName@bizlinktech.com). Please return the signed copy of the FSN to BizLink (bfr_vigilance_orion@bizlinktech.com) within 10 working days
- Root cause
- Software design
- Status
- Open, Classified
- Product code
- JAI
- Quantity affected
- 1 unit
- Distribution
- US Nationwide distribution in the states of MA, TN.
- Date initiated
- 2023-10-06
- Date posted
- 2023-11-21
- Location
- Chartres, France
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| LEONI ORION System (K160518) | Leoni Cia Cable System | 2016-07-14 |