Atlas FDA

Optical Access System; 5mm x 100mm and 12mm x 100mm; Model #: CTR03 and CTR73. Indicated for use to establish a port of

FDA device recall Z-1835-2015 · posted 2015-06-25 · Terminated

Recall details

Recalling firm
Stryker Sustainability Solutions
Reason for recall
Stryker Sustainability Solutions is recalling Trocars because they have received reports indicating that some may fracture at the distal tip resulting in fragments at the surgical site.
Action taken
On 06/01/15 a customer notification letter was sent out by the firm. The notification letter instructs customers to stop using the affected lots of the recalled product, and should consider other trocar models reprocesses by Stryker as an alternative. The letter notes that attached is a Effectiveness Check Form and complete list of affected lots shipped to the facility. A Stryker Sustainability Solutions Sales Representative will be sent to check the inventory of the affected lots. Once the inventory is checked the affected lots will be removed. The firm will complete a Recall Effectiveness Check Form . The form is to be completed even if no affected product is found. The firm requests that the form be returned to the sales representative or fax to 480-809-4975. If the form indicates that affected devices remain in inventory a prepaid shipping label will be issued for the return of the product. Customers will receive a credit for all affected devices returned. The firm states that adverse reactions or quality problems experiences with the use of the product may be reported to the firm's hotline 888-888-3433 x5572, or at www.stryker.com/productexperience as well as through FDA's MedWatch Adverse Event Reporting program.
Root cause
Under Investigation by firm
Status
Terminated
Product code
NLM
Quantity affected
147,351
Distribution
Distributed Nationwide including Puerto Rico.
Date initiated
2015-06-01
Date posted
2015-06-25
Date terminated
2016-08-16
Location
Tempe, AZ

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
REPROCESSED ENDOSCOPIC TROCARS AND CANNULAS (K063788)Ascent Healthcare Solutions2007-10-15