OPMI VARIO 700 Surgical Microscope. Manufactured by: Carl Zeiss Meditec AG, Oberkochen, Germany. Provide a magnified vie

FDA device recall Z-0105-2012 · posted 2011-10-24 · Terminated

Recall details

Recalling firm
Carl Zeiss Meditec, Inc.
Reason for recall
An incorrect nut and bolt set was used in the axis three (3) spring assembly of the OPMI VARIO 700 surgical microscope distributed to three US locations.
Action taken
Carl Zeiss Meditec had Field Service Engineers directly contact all affected customers. The Field Service Engineers will replace the defective component. Questions regarding this Field Corrective Action should be directed to the Director of Clinical/Regulatory Affairs at 925-557-4616.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
EPT
Quantity affected
3
Distribution
USA (nationwide) including the states of FL, GA, and LA .
Date initiated
2011-09-26
Date posted
2011-10-24
Date terminated
2011-11-04
Location
Dublin, CA