Omron brand 3-Way Instant Thermometer, Model Numbers MC-600 (USA) and MC-600CAN (Canada); an OTC device used to take an

FDA device recall Z-0466-06 · posted 2006-02-04 · Terminated

Recall details

Recalling firm
Omron Healthcare, Inc.
Reason for recall
Omron is recalling the Omron 3-Way Instant Thermometers due to possible overheating of the probe tip.
Action taken
Omron sent recall letters dated 10/21/05 to all of their direct accounts, informing them that the probe tip could overheat and possibly burn the person whose temperature is being taken. The accounts were instructed to check the lot number on their thermometers immediately to see if they fall within the range of affected thermometers (lots beginning with 01-32, 01-36, 01-37 and 01-38). The accounts were instructed to call Omron at 1-800-634-4350 if their unit is affected to get instructions on how to return the unit to Omron for a refund. Omron issued a press release on 10/24/05 to alert consumers who may have purchased the thermometer at a retail pharmacy, and has posted both the press release and the recall letter on their website at www.omronhealthcare.com.
Root cause
Other
Status
Terminated
Product code
FLL
Quantity affected
Approximately 48,000 units
Distribution
Nationwide and internationally to Canada, Chile, Venezuela and Singapore
Date initiated
2005-10-25
Date posted
2006-02-04
Date terminated
2007-11-28
Location
Bannockburn, IL