Omnipod 5 Automated Insulin Delivery System, iOS Application. Model Number: PT-000664/M009-S-AP. Software Version: all P
FDA device recall Z-2016-2025 · posted 2025-06-24 · Open, Classified
Recall details
- Recalling firm
- Insulet Corporation
- Reason for recall
- Insulet Corporation is recalling Omnipod 5 iOS App, due to a software design issue in which a comma is not recognized as a decimal separator and the user cannot enter a decimal separator in any manual entry screen if the user manually updates their phone settings to a region OUS that uses a decimal separator instead of a period decimal separator or chooses that format as a preference. Use of the affected software may result in over delivery of insulin by increasing the amount of insulin when a decimal separator is not recognized (e.g., 0.1 unit recognized as 1 unit) and thus may result in hypoglycemia, potentially severe.
- Action taken
- Insulet notified health care professionals and users on 05/23/2025 via emailed letter. Consignees were instructed to make sure your phone s keyboard and region settings match the approved country for use currently, the U.S. only and this issue does not affect the Insulet-provided Omnipod 5 Controllers or the Omnipod 5 App on compatible Android phones. A software update will be pushed to all devices in the coming weeks, and each consignee will be notified via email and in-app notification on the device when it is ready to install. The fix will allow commas to be recognized as decimal points when using the Omnipod 5 App for iPhone. The email notification to users included a section to formally acknowledge the notification.
- Root cause
- Software design
- Status
- Open, Classified
- Product code
- QRX
- Quantity affected
- 122,823 downloads
- Distribution
- US market only- Although there are 122823 downloads of the application, our data shows there are only ~80k active users.
- Date initiated
- 2025-05-23
- Date posted
- 2025-06-24
- Location
- Acton, MA
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|