Omega IV and Omega V tables on GE Innova 2000 fluoroscopic x-ray system

FDA device recall Z-0620-2010 · posted 2010-01-13 · Terminated

Recall details

Recalling firm
GE Medical Systems, LLC
Reason for recall
Potential table rotation of Omega IV and Omega V tables affecting Advantx-E and Innova 2000, 2100IQ, 3100, 3100IQ, 4100, 4100IQ, 2121IQ and 3131IQ cardiovascular X-ray imaging systems. Due to potential weakness of some table rotational brakes, an unintended table rotation may occur while the table is locked. This issue could lead to a patient fall if the unintended motion occurs during patient
Action taken
Consignees were sent on 12/22/09 a GE Healthcare " Urgent Medical Device Correction" letter dated 12/1/09. The letter was addressed to Hospital Administrators / Risk Mangers, Mangers of Radiology / Cardiology, Radiologists / Cardiologists. The letter described the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information.
Root cause
Device Design
Status
Terminated
Product code
MQB
Quantity affected
865
Distribution
Worldwide distribution: USA, SRI LANKA, ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, BELARUS, BELGIUM, BOLIVIA, BOSNIA & Herzegovina, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CZECH REPUBL
Date initiated
2009-12-22
Date posted
2010-01-13
Date terminated
2015-11-13
Location
Waukesha, WI

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
INNOVA 2000 AND INNOVA 2000S (MOBILE VERSION) (K024200)Ge Medical Systems, Inc.2003-01-16
INNOVA AND INNOVA 2000 S (K022322)Ge Medical Systems, Inc.2002-10-08
LCV+ VERSION 2 (K993037)Ge Medical Systems, Inc.2000-02-14