Olympus POWERSEAL 5MM, 44CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0544CJDA. An electrosurgical de

FDA device recall Z-1815-2022 · posted 2022-09-26 · Open, Classified

Recall details

Recalling firm
Olympus Corporation of the Americas
Reason for recall
An out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment
Action taken
Olympus issued Urgent Medical Device Recall Letter on 8/12/22 to Surgical Department Risk Management. Letter states reason for recall, health risk and action to take: 1. Immediately locate any product with affected lot numbers listed in this communication. See Attachment 2 for pictures showing the location of lot numbers on the package. 2. Call your Olympus customer service representative at 1 800 848 9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge to you. Olympus will issue a credit to your facility upon return of affected product. 3. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number 0412 and provide your contact information as indicated in the portal. 4. If you have distributed these devices outside your facility, please notify your customers of this removal action immediately. Your notification to your customers may be enhanced by including a copy of this recall notification letter. Contact directly at (647) 999-3203 or at Cynthia.Ow@Olympus.com for any additional information or support concerning this matter.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
GEI
Quantity affected
44 units (5/box); 133 units (5/box) OUS
Distribution
Nationwide OUS: AUSTRALIA, CANADA , HONG KONG, CHINA, KOREA
Date initiated
2022-08-12
Date posted
2022-09-26
Location
Center Valley, PA