Olympus High Flow Insufflation Unit. Model/ Number: UHI-2. Used for insufflation of the abdominal cavity, and automatic
FDA device recall Z-1353-2026 · posted 2026-02-18 · Open, Classified
Recall details
- Recalling firm
- Olympus Corporation of the Americas
- Reason for recall
- Issue with software algorithm which may lead to overpressure events.
- Action taken
- Olympus notified consignees on about 01/16/2026 via "URGENT: MEDICAL DEVICE REMOVAL" letter. Consignees were instructed to cease usage of the affected product immediately. They are advised to use available alternatives including an existing UHI-4 if available or switching to an insufflation device from another manufacturer. If no alternative is available on site, contact Olympus to arrange a loaner or purchase a replacement. Additionally, consignees were instructed to examine inventory and quarantine any identified devices, immediately cease usage of any UHI, UHI-2, and/or UHI-3 in inventory, call Olympus Customer Service at 1-800-848-9024, option 2 to obtain a Return Material Authorization to return your device. Consignees were also requested to acknowledge receipt of this letter through their recall web portal as follows: a. Go to https://olympusamerica.com/recall, b. Enter the recall number "0489 and c. Complete the form as instructed. Please indicate in your response if you no longer have the device(s). If any consignees have further distributed this UHI product, they were instructed to identify affected customers and forward this notice to them immediately. If you require additional information, please do not hesitate to contact me by phone at (647) 999-3203 or by email at Cynthia.Ow@Olympus.com.
- Root cause
- Software design
- Status
- Open, Classified
- Product code
- HIF
- Quantity affected
- 744 units
- Distribution
- Nationwide distribution
- Date initiated
- 2026-01-16
- Date posted
- 2026-02-18
- Location
- Center Valley, PA