Atlas FDA

Olympus Endoscopic Flushing Pump, Model OPF-2. Facilitates washing of tissues to remove blood, feces and other organic m

FDA device recall Z-0329-2014 · posted 2013-11-15 · Terminated

Recall details

Recalling firm
Olympus America Inc.
Reason for recall
There is a possibility that the pump head was not shipped in the locked position. If the pump head is in an unlocked position this may, under certain circumstances, cause the pump to stop and go into power stand-by mode.
Action taken
Notification letters were sent to customers on May 20, 2013.
Root cause
Component design/selection
Status
Terminated
Product code
FEQ
Quantity affected
263
Distribution
Worldwide distribution: USA (nationwide) and the country of Canada.
Date initiated
2013-05-20
Date posted
2013-11-15
Date terminated
2016-10-24
Location
Center Valley, PA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
OLYMPUS ENDOSCOPIC FLUSHING PUMP, MODEL OPF-2 (K100899)Keymed (Medical & Industrial Equipment), Ltd.2010-09-13