Olympus ENDOEYE HD II Rigid Videoscope Catalog Numbers: WA50040A, WA50042A -used with a video system center, light sourc
FDA device recall Z-0697-2024 · posted 2024-01-10 · Open, Classified
Recall details
- Recalling firm
- Olympus Corporation of the Americas
- Reason for recall
- Reports of a pink or green coloration of the image. If the issue is encountered during a procedure, device exchange could potentially result in prolonged surgery; potential visual impacts mucosal injury or bleeding may occur.
- Action taken
- UPDATE: On 2/9/24 Olympus issued Correction to original letter dated November 6, 2023, regarding the IFU references in #2. The purpose of this Field Corrective Action is to reinforce the requirement in the IFU that users inspect the image prior to a clinical procedure (as described in Section 4.3.3 and 4.4), always have a spare laparoscope available (as described in Section 2.5 General Safety Notes) and to check the image for the spare equipment. Olympus issued Urgent Medical Device Correction letter on November 6, 2023 to US Operating Room Manager, Risk Management Department. Letter states reason for recall, health risk and action to take: This Field Corrective Action reminds customers to follow the steps in the instructions for use, especially to inspect the image prior to a clinical procedure, and always have a spare laparoscope available. Olympus is currently investigating technical solutions to address this issue. Carefully read the content of this notification. 2. Please make sure that all medical personnel are completely knowledgeable and thoroughly trained in the ENDOEYE HD II Instructions For Use (IFU). The purpose of this Field Corrective Action is to reinforce the requirement in the IFU that users inspect the image prior to a clinical procedure (as described in Section 7.4 Testing), always have a spare laparoscope available (as described in Section 2.5 General Dangers, Warnings and Cautions) and to check the image for the spare equipment. 3. If you have further distributed this product, identify your customers and forward them this Customer Notification. Please appropriately document your notification process and let us know the end-customer feedback accordingly. 4. Complete the enclosed response form and return it to our third-party partner, Sedgwick, via e-mail at olympus5225@sedgwick.com or by fax 866-551-2707. For any questions about the acknowledgement form, please call the Sedgwick team at 866-875-0524. If you require additional information o
- Root cause
- Component design/selection
- Status
- Open, Classified
- Product code
- HET
- Quantity affected
- WA50040A : 561 units; WA50042A: 857 units
- Distribution
- Worldwide - US Nationwide distribution.
- Date initiated
- 2023-11-06
- Date posted
- 2024-01-10
- Location
- Center Valley, PA